6 Set up with the display and adjustment module
VEGAPULS 6X • Modbus and Levelmaster protocol
1015562-EN-240311
Bulk solids:
2
1
3
100%
0%
1 m
(39.37
")
35 m
(1378
")
Fig. 10: Parameterisation example min./max. adjustment - bulk solids
1 Min.level=max.meas.distance(distanceB)
2 Max.level=min.meas.distance(distanceA)
3 Reference plane
If these values are not known, and adjustment can for example be
carried out with the distances of 10 % and 90 %.
Thestartingpointforthesedistancespecicationsisalwaystherefer-
enceplane,e.g.thesealingsurfaceofthethreadorange.Informa-
tion on the reference plane can be found in the chapters "Mounting
instructions" resp. "Technical data".Theactualllingheightisthen
calculated on the basis of these entries.
The actual product level during this adjustment is not important,
because the min./max. adjustment is always carried out without
changing the product level. These settings can be made ahead of
time without the instrument having to be installed.
Proceed as follows:
1. Select with [->] the menu item Distance A (max. value) and con-
rmwith[OK].
2. Edit the distance value with [OK] and set the cursor to the re-
quested position with [->].
3. Adjust the requested distance value for 100 % with [+] and store
with [OK].
Distance A (max. value)