DGT / DGT60 / DGTQ / DGTP
50
6.6.3 MULTISCALE STRING
The string varies depending upon the configured channels:
Chan. Transmitted string
1 [CC]hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|"NO DATE TIME") + CR + LF
2 [CC]hh,pppppppp,uu,hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|"NO DATE TIME") + CR + LF
3 [CC]hh,pppppppp,uu,hh,pppppppp,uu,hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|"NO DATE TIME") +
CR + LF
4 [CC]hh,pppppppp,uu,hh,pppppppp,uu,hh,pppppppp,uu,hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|"NO
DATE TIME") + CR + LF
in which:
[CC] = INSTRUMENT CODE IN THE FORMAT OF TWO ASCII DECIMAL DIGITS
ONLY IF THE 485 PROTOCOL HAS BEEN SELECTED (FOR EXAMPLE 00)
For each set channel:
hh ST Stability of the display
US Instability of the display
VL Value in microvolts relative to the weight
RZ Value in converter points relative to the weight
, Comma character
pppppppp 8 digits (including eventual sign and decimal point) which identify the weight. The
insignificant digts are filled with spaces. Through the MVOL and RAZF commands
the indicator transmits the relative value on 10 digits instead of 8.
, Comma character
uu Unit of measure "kg" "bg" "bt" "lb" "mv" (microvolts) "vv" (converter points)
, Comma character (only with REXD command)
dd/mm/yy Date in the “dd/mm/yy” format (only with REXD command)
bb 2 space characters, ascii decimal 32 character (only with REXD command)
hh:mm:ss Time in the "hh:mm:ss" format (only with REXD command)
CR Carriage Return (ascii decimal 13 character)
LF Line Feed (ascii decimal 10 character)
If the optional TIME DATE board has not been detected (standard only in DGTP version), in response to the
REXD command, only the weight is transmitted and not the date and time; in its place there is "NO DATE
TIME".
LEGEND:
b space character, 32 decimal ascii character.