DGT / DGT60 / DGTQ / DGTP
64
10. ERROR MESSAGES
While using the indicator, it is possible to incur in the following errors:
MESSAGE DESCRIPTION
PREC. It is displayed if one tries to calibrate a point without first having confirmed the number of
calibration points
ERMOT Weight unstable during the acquisition of a point during calibration.
ERPNT During the acquisition of a calibration point a null value has been read by the converter.
Err.X.1: Error which takes place if the capacity of channel X has not been set, or if there is an error in
the calibration parameters of channel X, in which X indicates the number of channel to which
the error refers to.
Er – 11 Calibration error: a too small sample weight has been used; it is advisable to use a weight
equal to at least half of the scale capacity.
Er – 12 Calibration error: the acquired calibration point (tP1 o tP2 o tP3) is equal to the zero point
(tP0).
Er – 37 Scale to be calibrated TECH.MAN.REF. (it is advisable to carry out a technical default,
"dEFAu" parameter, before proceeding).
NOTE: press the T key before accessing the setup.
Er – 39 Scale to be calibrated TECH.MAN.REF. (it is advisable to carry out a technical default,
"dEFAu" parameter, before proceeding).
NOTE: press the T key before accessing the setup.
C.Er. – 36 During the calibration some internal negative points have been calculated:
- the calibration point is less than the zero point.
- the signal is negative (check the connections)
C.Er. – 37 During the calibration some internal points less than the minimum value have been
calculated:
- the calibration point is equal to the zero point.
- A capacity too high in relation to the division has been set.
hW-Err HARDWARE ERROR: software not compatible with the installed hardware; the hardware
expansion component is missing which allows the software to function.