3. SAFETY
3.1 Personal Safety Instructions
When using your HP Plus Tower, basic safety precautions should always be followed, including the following:
The quantity of dust in the ambient air can be very different from one place to another
We recommend you replace the dust lters every 4000 hours.
In some dusty areas, this maintenance needs to be performed more frequently.
Clogged dust lter may damage the device or results in a bad nitrogen quality.
The power supply main cable connector is the disconnect device in case of
emergency. It must be easily accessible at any time.
If the HP Plus Tower/Rack is used in a way not specied by VICI DBS, the
equipment warranty becomes void.
3.2 Instrument Safety Instructions
The HP Plus Tower/Rack is a nitrogen generator.
Do not let it run with the outlet open to the air in a low volume room.
The increase of the nitrogen (and decrease of oxygen) may cause human disease. It is the
responsibility of the user to check if the room is appropriate or if an O
2
measurement is
necessary.
The HP Plus Tower/Rack drains oxygen as waste.
Oxygen ratio in the air could change around the device in a low volume or not
sufcient ventilated room.
If fuel gases are present around, their explosive or ammable limits could be
affected. It is the responsibility of the user to check if the room is appropriate or
if an O
2
measurement is necessary.