3. SAFETY
3.1 Personal Safety Instructions
When using your HP Plus Tower, basic safety precautions should always be followed, including the following:
• The quantity of dust in the ambient air can be very different from one place to another
We recommend you replace the dust lters every 4000 hours.
• In some dusty areas, this maintenance needs to be performed more frequently.
• Clogged dust lter may damage the device or results in a bad nitrogen quality.
• The power supply main cable connector is the disconnect device in case of
emergency. It must be easily accessible at any time.
• If the HP Plus Tower/Rack is used in a way not specied by VICI DBS, the
equipment warranty becomes void.
3.2 Instrument Safety Instructions
• The HP Plus Tower/Rack is a nitrogen generator.
• Do not let it run with the outlet open to the air in a low volume room.
• The increase of the nitrogen (and decrease of oxygen) may cause human disease. It is the
responsibility of the user to check if the room is appropriate or if an O
2
measurement is
necessary.
• The HP Plus Tower/Rack drains oxygen as waste.
• Oxygen ratio in the air could change around the device in a low volume or not
sufcient ventilated room.
• If fuel gases are present around, their explosive or ammable limits could be
affected. It is the responsibility of the user to check if the room is appropriate or
if an O
2
measurement is necessary.