21
NM PLUS HYDROGEN GENERATOR
USER MANUAL
www.vicidbs.com
Version 1.06.0000
7.2.1. Congure Parameters
ITEM DESCRIPTION OPTIONS/RANGE DEFAULT
Set Clock Adjust the internal clock - -
Low Pressure Error* Percentage of total pressure that needs to
be reached for activating the low pressure
0-100% 30%
Pressure Rise Sets how fast the pressure must increase 0.01 - 4.2 barg/min
0.2 - 60 psig/min
0.09
14
Pressure Drop Delay Sets a delay in seconds to ignore a pressure
drop (override low pressure alarm)
2 - 10 min 2
Auto Start After powering the generator, it goes back
to the status before power shutdown. Exam-
ple:
1. System running, producing hydrogen and
valve is open
2. Power shutdown
3. Power restarted
4. System starts running and then opens
the valves automatically after reaching the
internal pressure
Yes/No No
Beeper Sets whether the audible signal is activated
in the event of an alarm
On/Off On
Cascading Mode Sets the cascading mode. (see cascading for
more details)
On/Off Off
N Gen H2 in Cascading Sets the number of generators in the cascad-
ing chain
1-10 1
ID Number Sets the ID-Number 1-10 1
Remote Start/Stop Mode Congures the remote Start/Stop function Start/Stop, Start only,
direct control
Start/Stop
Remote Relay Mode Congures the remote relay contacts Normally open (NO)
Normally closed (NC)
NC
Auto-Rell If set to “On”, the pre-level water alarm is
used to trigger an external pump or valve to
rell the water tank
On/Off Off
Auto-Rell Timeout Within this time the internal water tank
must be lled. if a timeout occurs the alarm
“check auto rell” appears.
5-600 seconds 30 sec.
Pre Alarms in List If set to Yes, the pre alarms are also shown in
the alarm log.
Yes/No No
Pressure Units Sets the desired unit for the pressure mea-
surement
bar/psi/kPa bar
Temp. Units Sets the desired unit for the temperature
measurement
°C and °F °C
Volume Units Sets the desired unit for the volume mea-
surement
scm (standard cubic
meters) OR scf
(standard cubic feet)
scm
* = If set at 100% generator never goes to External Pressure Alarm.