4
3.FEATURES
3.1. GENERAL FEATURES
3.1.1. Display: LCD.
3.1.2. Max. display: 6000 digits, and auto polarity display.
3.1.3. Clamp open max.28mm.
3.1.4. Over-range display: "OL".
3.1.5. Data hold.
3.1.6. Relative Value measurement.
3.1.7. Sampling rate: 3 times/sec.
3.1.8. The symbol " " appears when battery is low.
3.1.9. Continuity test: buzzer alarms when < (50±10) Ω.
3.1.10. Auto range or Manual range.
3.1.11. Auto power off.
3.1.12. Power consumption: approx.3mA.
3.1.13. Power source: 2 x 1.5V batteries (AAA).
3.1.14. Operating environment: (0-40) °C, relative humidity: <70%.
3.1.15. Dimension: 123Wx270Dx35H mm.
3.1.16. Weight: approx. 280g (including batteries).
3.1.17. Accessories: operation manual x 1pc, test leads x 1 pair,
temperature probe TP01 x 1 pair, 1,5V AAA batteries x 2pcs,
carrying bag x 1pc (optional).
3.2. SPECIFICATIONS
3.2.1. Accuracy: ±(a% of reading + digits). To ensure accuracy, the ambient temperature
should be (23±5)°C and the relative humidity shall be less than 70%. The accuracy is
guaranteed for one year since production date.
3.2.2. Temperature co-efficiency: 0.1 of specified accuracy/1°C (<18°C or >28°C).
3.2.3. DC Voltage (DCV)
Range Accuracy Resolution
600mV ±(1.0% of reading + 5) 0.l mV
6V
±(0.5% of reading + 5)
0,001 V
60V 0,01 V
600V 0,1 V
Input impedance: 10M Ω. Overload production: 1000Vrms.