General System Parameters
SETUP 1 (Inside SETUP Menu)
VT200/220 Technical Manual, Rev 1.00.01 48 Doc # TM-VT200/220-EN
To edit parameters in the SETUP 1 menu:
1. The current parameter number is shown on the display. Use PRINT (
) to scroll to
the parameter you need (see the following section for descriptions of parameters).
2. To edit the parameter value, use P.TARE ( ) to increment the current digit and
TARE (
) to move to the next digit.
3. When the display is showing the required value, press PRINT (
) to confirm your
selection.
5.2.2 SETUP 1 Parameters
Parameters 1.2, 1.3, 1.5 and 1.7 are not in use in this version of the
VT200/220, and so are not detailed in the table below.
Par. Description Values
1.1 Enable totalizer. YES enables the totalizer. NO disables it. 0=NO 1=YES
1.4 Date format. Sets the date format in all printing formats. 0=d-m-y 1=m-d-y
1.6 Enable battery-backed RAM. If the battery-backed RAM option is
installed, this parameter enables saving date, time and the
accumulated total in persistent memory.
You must enable this parameter to be able to set date and time.
0=NO 1=YES
1.8 Enable function menu.
YES specifies that users can access the function menu by pressing
TARE for 3 seconds.
NO specifies that users should not be able to access the function
menu.
0=NO 1=YES
1.t Auto power off. Time for the unit to switch off if the scale is zero. 00-99 minutes
00=Disabled
1.E Number of samples in counting. 5-99