28Page of 44 DT_0006 Issue 16
A Global Leader in Respiratory Solutions
Problem Fault
Symptoms:
Tests begin automatically.
Volume accumulates automatically
without the subject blowing.
Very small VC or FVC test displayed.
Possible Cause/
Solution: (In
probable order)
Device(orifusingremoteowhead,the
owhead)isnotstationaryatthestart
of test. Hold it steady until the ‘Ready to
Blow’ prompt appears.
Return to previous menu and re-enter
the test.
Problem Fault
Symptoms:
Cannot read screen.
Possible Cause/
Solution: (In
probable order)
Ensure the on/off button was pressed.
The batteries may be low, plug in USB
cable attached to a computer or replace
the batteries.
Electronics failure. Contact support.
Problem Fault
Symptoms:
Reversed or no volume measurements.
Possible Cause/
Solution: (In
probable order)
Ensureowheadisproperlyconnected/
seated on device (if using the remote
owheadalsocheckthatthetubingis
connected correctly).
Electronics failure. Contact support.
8.1. Software Check
Information about the device may be obtained from the About
option intheCongurationmenu.Thisinformationmaybe
usedwhenenquiriesaremadetoVitalographoraserviceagent.
8.2. Product Useful Life Checks
To ascertain whether the device has exceeded its useful life
Vitalographrecommendscheckingtheowheadandtherealtime
clock.