32Page of 44 DT_0006 Issue 16
A Global Leader in Respiratory Solutions
14. Technical Specication
Product Vitalograph micro, Model 6300
Flow Detection Principal Fleisch type pneumotachograph
Volume detection
Flow integration sampling @
100Hz
Maximum test duration 90 seconds
Maximum displayed
volume
10 L
Volume Accuracy ±2.5%
Flow Measurement Range
Max.owrate±960L/min(±16
L/s)
Min.owrate±1.2L/min(±0.02
L/s)
PEF Accuracy
±10% or ±10L/min of the reading
(ISO 23747:2015)
Back pressure
Less than 0.1kPa/L/sec @ 14L/
sec (ATS/ERS 2019)
Operating temperature
range
ISO 26782 limits: 17–35ºC
Design limits: 10–40ºC
Operating humidity range 30%–75%
Ambient pressure range 850hPa–1060hPa
Performance standards the
Vitalograph micro meets or
exceeds
ATS/ERS 2019, ISO 23747:2015 &
ISO 26782:2009
Electrical Safety Standard EN60601-1:2006+A1:2013
EMC Standard EN 60601-1-2:2015 + A1:2021
Home use Standard EN 60601-1-11:2015
Coexistence Standard ANSI C63.27-2017
QA/GMP standards
EN ISO 13485, FDA 21 CFR 820,
CMDR SOR/98-282 & JPAL