ENG
36Page of 44 DT_0006 Issue 16
A Global Leader in Respiratory Solutions
RF emissions
CISPR 11
Class B
The Model 6300 micro
is suitable for use in all
establishments, including
domestic establishments
and those connected to the
public mains network (e.g. at
homeanddoctor’socesin
residential areas)
EN 60601-1-2:2015 + A1:2021 -Immunity tests
Immunity test Test level
Compliance level
Reached
Electrostatic discharge
(ESD)
EN 61000-4-2
Contact: ± 8 kV
Air: ± 2 kV, ±
4 kV,
± 8kV, ± 15 kV
Contact: ± 8 kV
Air: ± 2 kV, ± 4 kV,
± 8kV, ± 15 kV
Radiated RF
EN 61000-4-3
10 V/m
80MHz to
2700MHz
3V
2700 to
6000MHz
10 V/m
80MHz to 2700
MHz
3V
2700 to 6000MHz
Proximityeldsfrom
RF devices
EN 61000-4-3
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2