Arguments for 930i measurement:
<LO_OHM>, Selects LO_OHM function
<test current>, 3 - 30 A (40 amps with option 40A)
<frequency>, 50 - 100 Hz
<max resistance>, 1 microohm - 0.999 Ohms
<min resistance>, 0 – 0.999 ohms
<test time>; 1 - 999 S
Example:
10 ! File: T2.IBW
20 ! This test sends a LO_OHM command to the 944i, waits
30 ! for test completion, and then returns the results
50 DIM Result$[120]
60 ASSIGN @V944 TO 1410
70 REMOTE @V944
80 ABORT 14
90 CLEAR 14
100 ! Issue test and command to run.
110 OUTPUT @V944;"ADD,LO_OHM,2,0,5;RUN;" !For 944i
110 OUTPUT @V944;"ADD,LO_OHM,25,60,1.0E-1,0,5;RUN;" !For 930i
120 ! Poll until test is complete.
130 Stb=SPOLL(@V944)
140 IF Stb<8 THEN 130
150 PRINT "Stb=",Stb
160 ! Get test results.
170 OUTPUT @V944;"RSLT?;"
180 ENTER @V944;Result$
190 PRINT Result$
200 PRINT
210 BEEP
220 STOP
230 END