48
Specifications
AC DIELECTRIC MODE
AC Voltage Output:
Voltage uncertainty is equal to the square root of the selected output.
Examples:
Output Formula Accuracy
100 VAC Sqrt(100) ± 10.0 VAC
500 VAC Sqrt(500) ± 22.4 VAC
1000 VAC Sqrt(1000) ± 31.6 VAC
5000 VAC Sqrt(5000) ± 70.7 VAC
Output Frequency: 50Hz to 100Hz with 1Hz resolution independent of line power
frequency.
Output Waveform: Digitally synthesized, low distortion sinewave.
AC Current Sensing (AC Leakage):
Range Resolution Accuracy
13.00uA 10nA 0.5% of reading ± 1uA
1.300mA 1uA 0.5% of reading ± 10uA
40.0mA 100uA 0.5% of reading ± 500uA
Current Sensing: Resistive current (in-phase), reactive current (out-of-phase) or
total current.
Voltage Coefficient: 100pA/V
AC Current Limits:
Range Current Limit
100 to 5,000 VAC 35mA
AC Current Limits with option FR-4:
Range Current Limit
100 to 500 VAC 20mA
500 to 3500 VAC 35mA
3500 to 4000 VAC 30mA
4000 to 4200 VAC 25mA
4200 to 4400 VAC 20mA
4400 to 4600 VAC 15mA
4600 to 5000 VAC 10mA
DC DIELECTRIC MODE
DC Voltage Output: