6. ERRORS AND CORRECTIVE ACTIONS
General errors
1 - Missing signal "mains zero", entry 0
2 - Missing signal "mains zero", entry 1
3 - Mains frequency is not 50 Hz or 60 Hz.
Errors concerning the heating
15 - During running of a program: when the
temperature is increasing the needed rate of
increasing cannot be achieved.
19 - The required temperature is not reached for the
selected time.
20 - The required temperature can not be reached
during cooling.
21 - The selected temperature can not be reached
during heating.
22 - Temperature is higher than 1200°С.
23 - Temperature is lower than 5°С.
99 - Interrupted thermo sensor or the temperature is
higher than 1285°С.
Errors concerning the movement of the table
54 - The sensor for the table to position drying is not
reached for a preset time.
76 - The time for opening is over during the program
controlled mode.
84 - The upper position sensor of the table (closing)
is not reached for the required time.
86 - The lower position sensor for the table is not
reached for the required time during opening of
the furnace.
ATTENTION!
Errors concerning the vacuum.
28 - The selected vacuum is not reached for the
required time.
50 - There exists residual vacuum during the initial
power on procedure.
51 - After releasing the vacuum, for the required
time the vacuum is not released entirely.
75 - The vacuum can not be released in programming
mode.
Errors concerning RS232 interface to the PC.
40 - A non-number-type symbol has appeared during
receiving.
41, 42 - A symbol is received which is not = or ~ during
receiving.
160 - Received number for the minutes greater than
59 during receiving.
161 - Received number for the seconds greater than
59 during receiving.
30, 31 - Timeout during receiving of programs.
121 - fault in the value of DRYTEMP
125 - fault in the value of FIRERATE
127 - fault in the value of IDLETEMP
128 - fault in the value of FIRETEMP
131 - fault in the value of VRTEMP
132 - fault in the value of VSTEMP
135 - fault in the value of FIRETIME
When any error appears, it is written on the display and
the furnace produces a sound signal. The button STOP
has to be pressed in order to exit from this state.