V
X
-
100 User Manual(v01
)
2
Check the following before use;
*Troubleshooting
In instances of abnormal operation, error messages will be displayed on the Control Panel.
If a problem persists, please request assistance from the customer support information services.
E ? : An error occurred. Turn the power off, and then turn it back on. If the error persists,
contact your Service Representative.
Error Code Check
Paramete
r
Description
E1 X-ray Generator In the "E1", "E2" and "E9" states where power is
maintained, the errors related to X-ray exposure
cannot expose the X-ray.
If the equipment is turned off and on after an error
related to X-ray exposure occurs, X-ray exposure is
performed normally.
E2
E9
BATT Battery Appears when the battery voltage is lower than the
reference value during X-ray exposure. Charge the
batter
y
.
CHG Appears when foreign matter or temporary
phenomenon in the charging hole of the equipment.
Check for foreign substances and connect and
disconnect the charger.
If it continues to connect and disconnect the charger
more than 5 times, it will be performed normally.
If the error code continues to occur, contact your
service representative.
Appears when the battery is charging. After
disconnecting the battery charger from the system,
start the s
y
stem.
BATT Appears when the battery level is lower than the
reference value during X-ray exposure.
A problem occurred. Turn the power off, and then turn it back on. If the problem persists,
contact your Service Representative.
Problem Cause Solution
Equipment is not turned on. The power switch is not
turned on properly.
Turn the equipment power switch off and
then on again.
Battery discharge Charge the battery with the charger and
check again.
Battery cable is not
connected properly.
Please contact your service
representative.
Battery defect Please contact your service
representative.
Control Panel is not turned on. Defective Control Panel
board
Please contact your service
representative.
Internal cable
disconnection
Please contact your service
representative.