100 User Manual(v01
)
3
No X-ray emission Exposure Switch Please contact your service
representative.
Internal cable
disconnection
Please contact your service
representative.
Defective generator Please contact your service
representative.
X-ray emission works, but
exposure is too light or
completely white.
The device is
positioned incorrectly.
Adjust the position of the equipment.
The exposure time is
too lon
.
Reduce the setup time.
Receptor direction
problem
Adjust the direction of receptor.
X-ray emission works, but
exposure is too dark.
The exposure time is
too short.
Increase the setting time.
* Checking for battery
1. Environment requirements for battery
(1) Storage Temperature
* Under 1 month : -10 - +45 ℃
* Under 3 month : -10 - +30 ℃
* 1 year : 25 ± 3℃
* Best Temperature : 23 ± 3℃
(2) Operating Temperature
* Charging : +10 - +45℃
* Discharging : -10 - +50℃
(3) Humidity : < 75% RH
2. Checking period for battery(Battery is a consumable component.)
-Recharge the battery every 3 months with checking the battery recharging condition.
-Recharging capacity shall be about 80%(3 level on the battery recharging indicator of control
panel) for storage.
-Recharge the battery with provided adapter within 30 mins for 80%(3 level on the battery
recharging indicator of control panel) recharging.
-Do not use the battery with liquid spillage or soaked.
-Dispose the battery according to a national law.