www.vtiinstruments.com
EX1200 Series Maintenance 143
FIGURE 9-2: EX1267/68/69 DMM REMOVAL
NOTE Once the DMM is removed from the mainframe, the chassis can no longer be used.
Installation Procedure
1) Using the guides shown in Figure 9-2, slide the DMM into the mainframe unit the until DMM
is supported by the card guides.
2) Insert the connector removed in step 2 above and insert it into P2, P3, and P4.
3) Slide the DMM into the mainframe unit the front panel is flush with the front of the chassis.
4) Install the three screws as shown in Figure 9-2 using a #2 Phillips screw driver.
5) Make sure the cables assemblies are inserted correctly, refer to Figure 9-3 for correct
orientation.
FIGURE 9-3: DMM CABLE INSTALLATION, CONNECTOR ORIENTATION

Table of Contents