PImax (MIP) Test
Slide the MicroRPM switch from ‘O’ to ‘MIP/MEP’, whilst applying
no pressure to the mouthpiece. Rotating segments will be displayed
whilst the unit performs an auto-zero function.
When the MicroRPM is ready a ‘beep’ will be heard and ‘0’ displayed.
Operation – Mouth Pressures (PImax/MIP +
PEmax/MEP)
Insert the battery into the compartment at the rear of the
MicroRPM.
Fit the required Pressure Valve Assembly (‘Inspiratory’ for PImax
(MIP), ‘Expiratory’ for PEmax (MEP)) into the MicroRPM; insert a
new Bacterial Filter into the Pressure Valve Assembly and then
the Rubber Flanged Mouthpiece onto the Bacterial Filter, as shown
below.