Electromagnetic Compatibility (EMC)
to IEC 60601-1-2
WARNING: Use of portable phones or other radio frequency
(RF) emitting equipment near the system (< 30cm) may cause
unexpected or adverse operation
The MicroRPM has been tested to IEC 60601-1-2:2014, regarding its ability
to operate in an environment containing other electrical/electronic equipment
(including other medical devices).
The purpose of this testing is to ensure that the MicroRPM is not likely to
adversely aect the normal operation of other such equipment and that other
such equipment is not likely to adversely aect the normal operation of the
MicroRPM.
Despite the testing of the MicroRPM that has been undertaken, normal
operation of the MicroRPM can be aected by other electrical/electronic
equipment and portable and mobile RF communications equipment. Keep a
distance of about 2 meters from possible error sources when using the device.
As the MicroRPM is medical equipment, special precautions are needed
regarding EMC (electromagnetic compatibility).
It is important that the MicroRPM is congured and installed/put into service,
in accordance with the instructions/guidance provided herein and is used only
in the conguration as supplied.
Changes or modications to the MicroRPM may result in increased emissions
or decreased immunity of the MicroRPM in relation to EMC performance.
The MicroRPM should be used only with the accessories (RS232 cables)
supplied (which are referenced in the accessories section of this manual).
None of the MicroRPM cables should be extended in length by the user.
If any cables are extended by the user or non approved accessories are
used, this may result in an increased level of emissions or decreased level of
immunity, in relation to the MicroRPM EMC. None of the MicroRPM accessories
should be used with other devices, as this may result in an increased level
of emissions or decreased level of immunity, in relation to the other devices’
EMC.
The MicroRPM has an essential performance – The respiratory pressure
readings on the product must remain within a tolerance of +/- 3% and the
unit rmware must not cease responding. Warning: In the event the product
is operated in the presence of signicant Electromagnetic Fields (particularly in
the frequency range 40-60MHz), whilst in the PC connected mode, ensure the
results on the unit and PC are the same. If the results dier then re-locating
the product away from sources of interference should resolve any issue.
List of EMC important cables. Do not exchange by other types
Part Number Description
36-CAB1000 Interface Cable for PUMA Software