Calibration
The calibration is factory set and should remain stable indenitely.
The calibration may be checked and altered, however, by connecting
the device via the Pressure Calibration Kit (36-ASS1234) to a Water
Manometer, as shown below:
Slide the MicroRPM switch from ‘O’ to ‘SNIP’ whilst applying no
pressure to the unit.
Gently ll the syringe until negative pressure of between 200 and
300cmH2O is read on the Water Manometer. Check that the reading
on the MicroRPM is within 3% of this reading.
PeraSafe (36-SSC5000A). Rinse thoroughly and leave to dry before
reassembly.
Important note: Nasal Probes, which are not sterilized, must be
disposed o immediately after each use.
If there are changes on the material surfaces (cracks, brittleness)
the respective parts must be disposed o.