Please Note: Calibration can only be adjusted in a positive
direction, as the MicroRPM measures the highest peak value.
Therefore, if the reading was greater than that on the Water
Manometer, the calibration screw must be turned anti-clockwise
rst, before calibration is attempted.
To calibrate the MicroRPM, connect to the Water Manometer as
before, lling the syringe to obtain the required negative pressure.
To change the calibration, slowly turn the calibration screw in
a clockwise direction until the same value is displayed on the
MicroRPM.
Servicing
If your unit requires service or repair please see page 144 for
contact details.
A full service manual including circuit diagrams and parts list is
available on request.