Product Lifetime
The MicroRPM is designed for a product lifetime of 5 years.
Trouble Shooting Information
Should you encounter problems operating your MicroRPM unit,
please consult the table below:
Problem Possible Cause Solution
unit will not turn on Battery is exhausted Replace the
battery
Slide switch
connetion
Return unit for
servicing.
Display shows reading
before test has been
performed
Internal tubing to pressure sensor
kinked
Return unit for servicing
21