WARNING: The MicroRPM should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use with other equipment is
necessary, the MicroRPM and the other equipment should be observed /
monitored, to verify normal operation in the conguration in which it will
be used.
The MicroRPM has an essential performance – The respiratory pressure
readings on the product must remain within a tolerance of +/- 3% and the
unit rmware must not cease responding.
Note: In the event the product is operated in the presence of signicant
Electromagnetic Fields (particularly in the frequency range 40-60MHz),
whilst in the PC connected mode, ensure the results on the unit and PC
are the same. If the results dier then re-locating the product aw
ay from
sources of interference should resolve any issue.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Micro RPM is intended for use in the electromagnetic environment specied
below. The customer or the user of the Micro RPM should assure that it is used in
such an environment
Emission Test Compliance Electromagnetic Enviroment -Guidance
RF emissions
CISPR 11
Group 1 The Micro RPM uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
RF emissions
CISPR 11
Class B The Micro RPM is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes
Harmonic emissions
IEC61000-3-2
Not Applicable
Voltage uctuations/
icker emissions
IEC61000-3-3
Not Applicable