Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
The Micro RPM is intended for use in the electromagnetic environment
specied below. The customer or the user of the Micro RPM should assure
that it is used in such an environment.
Immunity Test IEC 60601 test
level
Compliance
Level
Electromagnetic
Environment
-Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
Floors should be
wood, concrete
or ceramic tile. If
oors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
100 kHz
repetition
frequency for
power supply
lines
± 1 kV
100 kHz
repetition
frequency for
signal lines
Not Applicable Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Surge
IEC 61000-4-5
± 0,5 kV, ± 1 kV,
± 2 kV
Not Applicable Mains power
quality should be
that of a typical
commercial
or hospital
environment.
25