Maintenance program
MAINTENANCE PROGRAM TABLE
Frequency (*) # of cycles (*) Operation Consumable Performed by
Monthly 50
Clean the door seal and the chamber face side
See ANNEX 6
User
Clean the chamber, trays and the rack
Clean the chamber filter
Clean the external sterilizer surfaces
Clean the steam diffuser plate
3 months 400
Replace the bacteriological filter
Replace the dust filter
6 months 800 Clean both water tanks
Yearly 800 Replace the door seal
5 years 4000 General check and service
Service technician
- 20000 General check and service
(*) whichever occurs first