9
4. Installation and start-up
9
Required minimum clearances
Back side: 50 mm
Right and left sides: 10 mm
Upper side: As required for filling the water tank, 50 mm minimum
Placement
Place the sterilizer on a flat and level surface, far from sources of heat and from flammable materials.
Do not place the sterilizer so that it is difficult to open the service door and operate on the controls in it.
Do not place the sterilizer so that it is difficult to disconnect the power supply plug.
Place the sterilizer in a well ventilated room.
If installed in a cabinet, this shall be provided with an opening of at least 200x150 mm on the rear side.
The sterilizer must not be operated in presence of explosive atmospheres.
Always wear personal protective equipment.