15Andrew+ Robot User Manual
[ ANDREW+ ROBOT USER GUIDE ]
H. MAINTENANCE
i. General information
The Andrew+ Robot does not require periodic maintenance, if
used properly and in a clean environment. Nevertheless, it is
possible to carry out periodic checks and very simple actions
to guarantee the best status and results of the Andrew+
Robot.
ii. Recommended maintenance if the Andrew+ Robot
is moved
If the Andrew+ Robot is moved, we strongly suggest cleaning
the silicone mat on the bottom of the baseplate using acetone.
By doing so, the required gripping between the Andrew+
Robot and the bench is guaranteed. The bench to which the
Andrew+ Robot will be moved must be cleaned as well.
iii. Compatible solvents with the Andrew+ Robot
To clean all the parts of the Andrew+ Robot, the compatibility
with the following solvents have been tested:
■
Isopropanol alcohol
■
Ethanol
■
2% bleach solution.
iv. Pipette best practices
1. No stickers should be added on the pipettes sides to avoid
grabbing performances degradation.
2. Monthly electromagnets cleaning with alcohol/IPA is
recommended to keep good connectivity and avoid pipette
detection issues in the Andrew+ Robot’s hand (a guide could
be generated).
Use of soft tissue to clean all the Andrew+
Robot parts to avoid possible scratches on
their surfaces.
DO NOT clean the electronical connectors
present in the tool stand, waste Domino and
base plate.