No flow from fluidics on Xevo/SYNAPT/Vion - WKB50419
SYMPTOMS
• No beam from infusion
• With PEEK disconnected from source, no liquid flow is seen when trying to infuse from bottles
• Flow can be seen when infusing wash solution (after purging)
ENVIRONMENT
• Synapt HDMS
• SYNAPT G2
• SYNAPT G2 MS/HDMS
• SYNAPT G2-S MS/HDMS
• SYNAPT G2-Si MS/HDMS
• SYNAPT G2-S
• SYNAPT G2-Si
• Vion IMS QTof
• Xevo QTof
• Xevo G2 QTof
• Xevo G2 Tof
• Xevo G2-S
• Xevo G2-S QTof
• Xevo G2-S Tof
• Xevo G2-XS
• Xevo G2-XS QTof
• Xevo G2-XS Tof
• Xevo TQ MS
• Xevo TQ-S
• Xevo TQ XS
Waters - The Science of What's Possible.
https://support.waters.com/KB_Inst/Mass_Spectrometry/WKB50419_No_flow_from_fluidics_-_Xevo_Synapt_Vion
Updated: Sun, 08 Oct 2023 13:09:15 GMT
Powered by
1