EasyManua.ls Logo

Waters Prep Series - User Manual

Default Icon
198 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Waters Prep Series
Installation and Maintenance Guide
71500045204/Revision B
Copyright © Waters Corporation 2010
All rights reserved

Table of Contents

Questions and Answers

  • M
    Mrs. Kelli ShannonAug 16, 2025
    Why is there increased system backpressure during injection using Rheodyne injector in Waters Laboratory Equipment?
    • E
      Edgar RossAug 16, 2025
      Increased system backpressure during injection with a Rheodyne injector in Waters Laboratory Equipment is normal. When the injector is in the Inject position, the sample loop generates additional system backpressure. To decrease the system backpressure, return the injector to the LOAD position after loading the sample.
  • J
    Joseph MooreAug 19, 2025
    What to do if the Waters Prep Laboratory Equipment system controller screen is blank?
    • J
      Jonathan MckeeAug 19, 2025
      If the screen on your Waters Laboratory Equipment system controller is blank, the screen saver feature may be active. Press any function key to display the screen.
  • J
    John HarrisonAug 22, 2025
    What to do if my Waters Prep Laboratory Equipment pump keeps shutting down due to exceeding the high-pressure limit?
    • G
      Gregory RobertsonAug 22, 2025
      If your Waters Laboratory Equipment pump shuts down because the high-pressure limit is exceeded, there are several potential causes. It could be due to a high flow rate during injection valve actuation, causing a pressure spike; in this case, decrease the flow rate during injection. Another cause may be a blockage in the system after the transducer; try removing components between the pump and the end of the system to locate and eliminate the blockage. Finally, the column might be plugged, so try cleaning or replacing it.
  • A
    Angela ReedAug 25, 2025
    Why is my Waters Prep pump cavitating?
    • G
      George SnyderAug 26, 2025
      Pump cavitation in your Waters Laboratory Equipment can occur for several reasons. First, check the solvent reservoirs' position; if they are below the solvent proportioning valve, place them on the bench beside the fluid handling unit. Second, inspect tubing connections and fittings for leaks prior to the inlet check valve, as these can cause air aspiration; eliminate any leaks by tightening the fittings or replacing the ferrules. Lastly, the solvent inlet filter might be restricting flow, so clean or replace the filters.
  • S
    Stephen GutierrezAug 28, 2025
    What to do if Waters Prep Laboratory Equipment pump shuts down because the low-pressure limit is exceeded?
    • J
      jeannejacksonAug 28, 2025
      If your Waters Laboratory Equipment pump shuts down because the low-pressure limit has been exceeded, the issue is likely due to leaks, empty reservoirs, or bubbles in the pump head. Eliminate any leaks, fill the reservoirs, and prime the pump to resolve this.
  • C
    Christopher SullivanAug 31, 2025
    How to fix a grinding or shuddering noise from Waters Laboratory Equipment pump?
    • J
      jessicaharrisonAug 31, 2025
      If you notice a grinding or shuddering noise from your Waters Laboratory Equipment pump, and the high-pressure shutdown has not been exceeded, it could be due to a blockage in the system before the transducer. Clean the pump head outlet tubing, outlet check valve, and the tubing leading to the transducer. If the problem persists, it may be a mechanical issue.
  • E
    Edward ClarkSep 3, 2025
    Why is there a pulsing flow from the waste line of my Waters Prep Laboratory Equipment?
    • J
      Jeffrey BrewerSep 3, 2025
      A pulsing flow from the waste line in your Waters Laboratory Equipment is often caused by dirty check valve(s). Clean the check valve(s) to resolve this issue.
  • T
    Todd SanchezSep 6, 2025
    What to do if Waters Prep RI detector is leaking internally?
    • S
      Sara BradshawSep 6, 2025
      If your Waters Laboratory Equipment RI detector is leaking internally or from the purge port, the cell may be damaged because the maximum flow rate or pressure of the RI cell has been exceeded. Ensure you do not exceed 5 mL/min for 2414, 2410, and 410 Detectors, and do not exceed 100 psi for any Waters RI detector.
  • Z
    zmartinSep 9, 2025
    Why are peak retention times halved in Waters Prep?
    • C
      christopher65Sep 9, 2025
      Halved peak retention times in Waters Laboratory Equipment are caused by an incorrect pump head volume entered on the PUMP CONFIGURATION screen. For PrepLC operation with 1000-µL pump heads, enter 1000 for the pump head volume on the PUMP CONFIGURATION screen.
  • D
    Dana GrimesSep 12, 2025
    Why can't I enter a flow rate above 150 mL/min on my Waters Laboratory Equipment?
    • K
      Kelly SerranoSep 12, 2025
      If you are unable to enter a flow rate above 150 mL/min on your Waters Laboratory Equipment, it is because an incorrect pump head volume was entered on the PUMP CONFIGURATION screen. For operation with 1000-µL pump heads, enter 1000 for the pump head volume on the PUMP CONFIGURATION screen.

Summary

1 Overview

Waters Prep Series

Describes the features and configurations of the Waters Prep Series HPLC systems.

Getting started

Outlines the initial steps for using the system, including startup and basic operations.

Using the front panel

Guides on navigating and operating the instrument using its front panel controls.

Configuring the controller

Helps determine the appropriate controller configuration for system use.

2 Installing the System

Setting up and inspecting

Covers initial setup and inspection procedures for the system.

Electrical requirements

Details the electrical specifications and configurations for the system.

Assembling the system

Explains how to connect system components and modules.

Liquid line connections

Making tubing connections.

3 Troubleshooting, Testing, and Maintenance

Troubleshooting

Identifying and resolving system problems.

Performance tests

Tests to isolate problem causes.

Maintenance procedures

Routine maintenance tasks.

A Safety Advisories

Warning symbols

Explains hazard symbols.

Caution symbol

Explains caution symbol.

B Specifications

Operational specifications

Operational parameters.

Environmental specifications

Environmental ranges.

C Spare Parts

Common parts

Lists common spare parts.

150-mL configuration

Spare parts for 150-mL system.

D Solvent Considerations

Clean solvent

Importance of clean, particle-free solvents.

Solvent miscibility

Determining solvent miscibility.

Solvent degassing

Benefits of degassing solvents.

Waters Prep Series Specifications

General IconGeneral
Pressure RangeUp to 6000 psi
Detector OptionsUV/Vis, PDA, ELSD, MS
Detection Wavelength Range190 to 800 nm
Solvent DeliveryBinary or Quaternary
Fraction CollectionYes
Software ControlMassLynx, or Empower
Software CompatibilityWaters FractionLynx, MassLynx, or Empower
Pump TypePiston pump

Related product manuals