Making detector connections 2-49
goes to the detector’s signal input. Your detector type determines whether you
do this by interchanging two BNC signal cables or by keyboard input.
Using the RI detector with the flow splitter option
Tip: When using a Waters 2414 Detector with flow rates above 5 mL/min, use
the flow splitter option.
The flow splitter eliminates excessive pressure on the RI cell. It diverts about
1% of the flow to the RI detector and, ultimately, to waste. The remaining flow
goes to the function valve and fraction outlets or gets divided at the fraction
collector.
Tip: Consult the flow splitter option’s accompanying instructions for
installation details.