Troubleshooting 3-33
Program Method
Table Saved…
Programming: Keyboard is locked while the SAVE
function is storing the PROGRAM METHOD screen.
This message is accompanied by an audible alarm.
Check
Corresponding
Gradient
table/Events
table/Detector table
Programming: Since the Gradient, Event, and
Detector tables always run together (dictated by the
table number), modifications made to the Gradient
table should be consistent with the event or detector
table of the same table number. This message is
accompanied by an audible alarm.
An Inject Occurred Operation: Informational message indicating that an
inject signal was sent. This message is accompanied
by an audible alarm.
Line Is Incorrect Or
Incomplete
Programming: Finish the entries required before
attempting to move the cursor to a new location. This
message is accompanied by an audible alarm.
UV Detector 1 Not
Found
UV Detector 2 Not
Found
RI Detector Not
Found
Operation: Device has disappeared from the
IEEE-488 bus. Check if the PowerLine module was
turned on or if the PowerLine module was turned on
before the controller was turned on. Go to the
SYSTEM CONFIGURATION screen and rescan
IEEE-488 bus. This message is accompanied by an
audible alarm.
Illegal Vial Number Vial number is greater than the maximum value for
the selected carousel. This message is accompanied
by an audible alarm.
Number Out Of
Range
Check the range for the parameter using the HELP
key.
Composition MUST
sum to 100%
The composition values for %A, %B, %C, and %D do
not total 100%. Check entered values for all solvent
reservoirs. This message is accompanied by an
audible alarm.
Step Does Not Exist A step number was entered into the initial OPERATE
METHODS screen which does not exist in the method
table. Check the PROGRAM METHOD screen. This
message is accompanied by an audible alarm.
Operational warning messages (Continued)
Message Description