36 Using the ventilator
EN
No Assist Alarm
In “Assisted Ventilation” mode, patient fails to trig-
ger within the time window in two consecutive
phases.
The LED in alarm field flashes and the acoustic alarm
is triggered after a 1-minute delay.
Cancelling acoustic alarm
The acoustic alarm can be temporarily cancelled by
pressing alarm acknowledgement:
Stenosis: 30 Seconds
Disconnection: 30 Seconds
< 2,7 bar: 30 Seconds
: 120 Seconds
No Assist: 120 Seconds
The optical alarm will continue to flash.
If the cause of the alarm is not eliminated, the acous-
tic alarm will start to sound again after a short inter-
val.
Both the optical and the acoustic alarm are cancelled
automatically as soon as the cause is eliminated.