46 Functional checks
EN
Note Check the test bag before each functional
check. The balloon of the test bag must be un-
damaged and firmly connected to the connec-
tor. Have the test bag serviced at the same time
as the device.
7.2 Obligatory checks
Before each use:
Carry out a functional check.
After each use or dismantling:
Clean, disinfect or sterilize the ventilator and its
components (note “6. Hygienic preparation“ on
page 39);
•Check the:
lip membrane in the patient valve
valve membrane in the expiration arm
valve membrane in the spontaneous breath-
ing arm (note “7.4 Check of patient valve“
on page 48). They may be neither crinkled,
sticky or misshapen.
Carry out a functional check.
At least every 6 months, if the ventilator
has not been used in the intervening
period:
Carry out a functional check.