44 Hygienic preparation
6.6 Cleaning, disinfecting and sterilizing
The hygienic preparation should be performed on
MEDUMAT Standard a and the accessories used as
described in the following table.
Always follow the directions for use supplied with
the disinfectant. We recommend gigasept
®
FF (new)
for disinfecting by immersion and terralin
®
protect
for disinfecting by wiping. You are recommended to
wear suitable gloves (e.g. household or disposable
gloves) during disinfection procedures.
For further information on hygiene treatment and a
list of all cleaning agents and disinfectants which can
be used, please see our Internet brochure at
www.weinmann-emergency.com.
(1) After disinfection rinse the components thoroughly with dis-
tilled water and allow them to dry.
(2) Thermal disinfection in automatic cleansing device.
(3) Steam sterilization at 134 °C with devices to EN 285, dwell
time 5 minutes (or 18 minutes at 121 °C).
Description of
component
Cleaning Disinfection
Rinsing in
washing machine
Sterilization
MEDUMAT Standard a
With a dry or damp
cloth
Wiping Not permissible
Not
permissible
Patient valve
In warm water with
a mild household
detergent
Immerse in a weak
solution
(1)
Rinse program at
95 °C
(2)
Steam
sterilization at
up to 134 °C
(3)
Silicone ventilation
mask
Ventilation hose
Oxygen fittings
With a dry or damp
cloth
Wiping Not permissible
Not
permissible
Device outlet port filter
upgrade set
Follow the instructions for use for the device outlet port filter upgrade set
Hose casing, reusable
Wipe with a dry or
damp cloth
Rinse cycle 30°C,
no spin
Possible during
cycle
Not permitted