64 Servicing
EN
9.2 Sending in device
Warning!
Risk of infection from contaminated parts
during maintenance measures!
The device, components and accessories may be
contaminated and infect specialist staff with bacteria
or viruses when they are carrying out maintenance
measures.
Clean and disinfect device, components and ac-
cessories.
Do not send in potentially contaminated parts.
1. Remove components and accessories.
2. Clean the device, components and accessories
(note “6. Hygienic preparation“ on page 39).
3. Send device, and if necessary components and
accessories, to WEINMANN Emergency or to
specialist staff expressly authorized by
WEINMANN Emergency.
Note If you send in obviously contaminated parts,
these will be disposed of by WEINMANN Emer-
gency or by specialist staff expressly authorized
by WEINMANN Emergency at your expense.
9.3 Storage
If you do not intend to use MEDUMAT Standard a for
a long period, we recommend the following storage
precautions:
1. Clean and disinfect the ventilator (note “6.
Hygienic preparation“ on page 39).