Functional checks 45
7. Functional checks
MEDUMAT Standard a must not be used if the func-
tional checks reveal defects or deviations from the
selected parameters.
First try to correct the error with the help of the in-
formation provided in section “8. Troubleshooting“
on page 57. If this is not possible, have the device re-
paired by the manufacturer Weinmann or by special-
ists explicitly authorised to do so by same.
7.1 Preparation for functional check
For the functional check, you require:
– patient’s hose system
– test bag
– oxygen cylinder
– soap-and-water solution made from non-
perfumed soap
– adapter from test set WM 15357
We recommend that you hold reserve stocks of the
following items:
• washers for the connections;
• dust filters;
• lip membranes for the patient valve;
• membrane for spontaneous breathing arm;
• membrane for expiration arm.
1. Connect the device to the oxygen cylinder.
2. Connect the patient’s hose system to the device.