For cleaning the wallboard and VESA mount use only 70 percent isopropyl alcohol
solution applied to a clean cloth.
For the Braun ThermoScan PRO 6000 thermometer, use only the approved cleaning
agents published in the manufacturer's instructions for cleaning. Unapproved
cleaning agents can damage the device and interfere with data transmission.
Clean the Braun ThermoScan PRO 6000 contacts
Debris that accumulates on the Braun ThermoScan PRO 6000 electrical contacts can
interfere with data transmission. Welch Allyn recommends cleaning the contacts on the
thermometer and the dock once every 4 months to maintain optimal performance.
CAUTION Do not use bleach solutions of any kind when cleaning metal
electrical contacts. They will damage the device.
1. Slightly dampen a cotton swab with 70% isopropyl alcohol.
2. Remove the thermometer from the dock and clean the metal electrical contacts on
the thermometer with the cotton swab.
3. Place the thermometer aside for 1 minute, allowing the contacts to air dry.
4. Clean the metal electrical contacts on the device dock with the cotton swab.
Directions for use Maintenance and service 87