SureTemp® Plus Service Manual 15
3. Preventive Maintenance
The following preventive maintenance is recommended to maximize uninterrupted service
with the SureTemp
®
Plus Model 690 and 692 Thermometers.
3.1 Six-Month Maintenance
Units that are in service on a regular basis should have the following preventive maintenance
performed every six months:
1. Visually inspect the thermometer for any physical damage that might cause future product
failure.
2. Clean the unit per instructions in the Operator’s Manual supplied with the thermometer
and/or per the instructions below.
3.2 Twelve-Month Maintenance
For units in service on a regular basis, perform the Power-On Self Test and Model 9600 or
9600 Plus Calibration Testing procedure every 12 months. The preventive maintenance
procedure is located in the Model 9600 or 9600 Plus Operation Manual.
3.3 Standard Storage
Units which are stored for an extended period should have the following maintenance
performed every 12 months:
1. Remove the batteries before storage.
2. Reinstall the batteries and perform the Power-On Self Test and Model 9600 or 9600 Plus
Calibration Testing procedure found in the Model 9600 or 9600 Plus Operation Manual
before use.
3.4 Cleaning
3.4.1 Cleaning and Disinfection the Thermometer and Probe
Wipe the thermometer and probe with an appropriate health care low- or intermediate-level
cleaner/disinfecting wipe that incorporates either a 1:10 sodium hypochlorite (bleach) solution
or isopropyl alcohol as the active disinfection ingredient. Follow wipe manufacturer's
instructions for appropriate use, contact times and applicable warnings and precautions.
CAUTION: DO NOT immerse or soak the thermometer or probe in any type of fluid.
CAUTION: DO NOT use steam, heat, or gas sterilization on the thermometer or probe.
CAUTION: DO NOT autoclave the thermometer or the probe.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn SureTemp Plus 692 Series and is the answer not in the manual?

Welch Allyn SureTemp Plus 692 Series Specifications

General IconGeneral
BrandWelch Allyn
ModelSureTemp Plus 692 Series
CategoryThermometer
LanguageEnglish

Related product manuals