Electromagnetic Immunity
The Welch Allyn SureTemp Plus is intended for use in the electromagnetic environment specified below.
The customer or user of the Welch Allyn thermometer should assure that it is used in such an environ-
ment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Welch
Allyn SureTemp Plus models 690 and 692, including
cables, than the recommended separation distance
calculated from the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
Conducted
RF
IEC 61000-
4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = (1.17) P
Radiated RF
IEC 61000-
4-3
3 V/m
80 MHz to 1 GHz
3 V/m
d = (1.17) P 80 MHz to 800 MHz
d = (2.33) P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
a
should be
less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn SureTemp Plus 692 Series and is the answer not in the manual?

Welch Allyn SureTemp Plus 692 Series Specifications

General IconGeneral
BrandWelch Allyn
ModelSureTemp Plus 692 Series
CategoryThermometer
LanguageEnglish

Related product manuals