10 WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
4. Design and function
In addition to gauge and absolute pressure measurement, differential pressure can
also be measured with the CPH6200-S2 and two connected pressure sensors,
model CPT6200.
Selectable pressure units here are bar, mbar, psi, Pa, kPa, MPa, mmHg or inHg. An
integrated data logger and various other functions (such as Min., Max., Hold, tare, zero
point correction, alarm, power-off, variable measuring rate, sea level etc.) ensure that the
hand-held pressure indicator can be used for many different applications.
4.1 Display
1
Main display: Current measured value of sensor 1
2
Secondary display: Current measured value of sensor 2 or differential value
between sensor 1 and sensor 2
3
Logg arrow: Logger is ready
Arrow blinking: Automatic recording (Logg CYCL) active
4
Tare arrow: Tare function is activated
5
SL arrow: Altitude correction (sea level) is activated
6
Display arrows for measured value units
7
Indication elements for showing the Min./Max. measured value, difference or hold
6
6
7
1
2
45 3