29WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
6. Press the [SET/MENU] button.
Return to main menu.
6.5.2 Automatic recording with adjustable cycle time (Func CYCL)
The logger cycle time is adjustable (see configuration). As an example, with CYCL = 1:00
one measuring result will be stored every minute.
Additionally, with the measurement type rAtE SLo”, a current-saving function is
selectable, “Lo.Po”. If this is “on”, it operates so that, while the logger is recording,
a measurement only occurs at the respective logger time. This reduces the current
consumption considerably and is therefore mainly recommended for long-term
measurements (e.g. leak test).
Storable measuring results: CPH6200-S1: 10,000
CPH6200-S2: 4,000
(at max. 64 recording sequences)
Cycle time: 1 ... 3,600 s (= 1 h)
adjustable in the configuration
A measuring result contains:
Slow measurements “rAtE SLo”:
- Sensor 1: Measured value at that data point
- Sensor 1: Min. peak, max. peak since last data point
- Sensor 2
1)
: Measured value at that data point
- Sensor 2
1)
: Min. peak, max. peak since last data point
- Difference (sensor 1 - sensor 2)
1)
: measured value at that data point
- Difference (sensor 1 - sensor 2)
1)
: Min. peak, max. peak since last data point
Fast measurements “rAtE FASt or “rAtE P.dEt”:
- Sensor 1: Arithmetic mean value since last data point
- Sensor 1: Min. peak, max. peak since last data point
- Sensor 2
1)
: Arithmetic mean value since last data point
- Sensor 2
1)
: Min. peak, max. peak since last data point
- Difference (sensor 1 - sensor 2)
1)
: Arithmetical mean value since last data point
- Difference (sensor 1 - sensor 2)
1)
: Min. peak, max. peak since last data point
6. Commissioning, operation
1) Only valid for 2-channel version CPH6200-S2