28 WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
Sensor 2
1)
: Min. peak, max. peak since last data point
Difference (sensor 1 - sensor 2)
1)
: Measured value at that data point
Difference (sensor 1 - sensor 2)
1)
: Min. peak, Max. peak since last data point
Time and date stamp at that data point
With each recording, “St.XX” will be displayed briefly. XX represents the number of the
measuring result.
6.5.1.1 Delete stored data
1. Press the [STORE/QUIT] button for 2 seconds.
The main menu CLr. is accessed.
2. With the [▼/MIN] or [▲/MAX] button, select the desired function.
The following functions for deleting can be selected:
CLr
. no
Do not delete (cancel the process).
CLr
. ALL
Delete all data sets.
CLr
. LASt
Clear the last data set.
3. With the [STORE/QUIT] button, confirm the selection.
4. Press the [SET/MENU] button.
Return to main menu.
Logger memory is full
Lo66
FvLL
Shows in the event the logger memory is full. The recording will be
stopped automatically.
6.5.1.2 Reviewing individual values
In contrast to the cyclic logger function, individual values can also be viewed directly in
the display:
1. Press the [SET/MENU] button for 2 seconds.
Menu item “rEAd Lo66 will be accessed.
2. With the [▶/TARA] button, the last measuring result is displayed.
3. Press the [▶/TARA] button again.
Individual values of the measuring result are displayed.
4. With the [▼/MIN] or [▲/MAX] buttons, recall another measuring result.
5. Press the [▶/TARA] button.
Individual values of the new measuring result are displayed.
6. Commissioning, operation
1) Only valid for 2-channel version CPH6200-S2