EasyManua.ls Logo

WIKA CPH6200 - 10. Specifications; CPH6200 Digital Indicator Specifications

WIKA CPH6200
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38 WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
10. Specifications
10. Specifications
10.1 Digital indicator model CPH6200
Digital indicator model CPH6200
Electrical connection for reference pressure sensor
Measuring inputs
■
1 input for CPH6200-S1
■
2 inputs for CPH6200-S2
Sensor compatibility Compatible with reference pressure sensor model CPT6200
Connection to CPH6200 6-pin, shielded mini DIN female connector with interlocking
Sensor connection cable
■
Cable with 6-pin mini DIN connector and 7-pin bayonet
connector, length 1.1 m [3.3 ft]
■
Extension cable, length 3.8 m [12.5 ft], overall cable length
approx. 5 m [16.4 ft]
Display
Display indication Large 4 ½-digit LC display for indication of two pressure values
and additional information
Indication range -19999 ... 19999 digits (dependent upon connected reference
pressure sensor)
Pressure types Dependent upon connected reference pressure sensor
■
Gauge pressure, absolute pressure or vacuum
■
Differential pressure measurement only with CPH6200-S2, and
two model CPT6200 reference pressure sensors connected
Pressure units Freely adjustable depending on the measuring range
■
bar
■
mbar
■
psi
■
Pa
■
kPa
■
MPa
■
mmHg
■
inHg
Functions
Measuring rate Measuring rate (can be set via menu)
■
4/s (“Slo” - slow measurement)
■
1,000/s filtered (“FASt” - fast measurement)
■
> 1,000/s unfiltered (“P.det” - peak value detection)
Mean value filter 1 ... 120 seconds (can be set via menu)
Data logger
■
Individual value logger
⇒
Up to 99 recordings incl. time via button press
■
Cyclic logger
⇒
Automatic recording of up to 10,000 values incl. time
⇒
Cycle time freely adjustable in the range from
1 ... 3,600 seconds
Real-time clock For data logger, (can be set via menu)
Min./Max. memory Minimum or maximum measured value (can be accessed via
function button)

Table of Contents