33WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
Display Cause Measures
Err.1
Measured value above allowable
range
Check: Is the pressure over the
permissible measuring range of the
sensor?
Measured value too high!
Reduce pressure
Check the pressure measuring range
of the sensor and, if necessary,
replace with a suitable sensor with a
higher measuring range.
Sensor defect Send in for repair
Err.2
Measured value below allowable
range
Check: Is the pressure under the
permissible measuring range of the
sensor?
Measured value too low!
Reduce pressure
Check the pressure measuring range
of the sensor and, if necessary,
replace with a suitable sensor with a
lower measuring range.
Sensor defect Send in for repair
Err.3
Scale range exceeded Check: Is the value over 9999?
Value is too high!
Reduce value
Err.4
Under the scale range Check: Is the value below -2000
(tare?)
Value is too low!
Increase the value
Err.7
System error Send in for repair
Er.11
Measured value could not be
calculated
Choose different unit
Calculation range exceeded Choose different unit
7. Faults