EVOM™ Manual
World Precision Instruments 13
4. Press
Set Time
. Set the appropriate time and press
Return
.
5. Press
Time Format
. Choose 12 hour or 24 hour and press
Return
.
Fig. 19—(Left) Set the time and press Return
Fig. 20—(Right) Choose the time format and press Return
Probe Null
1. Press
Probe Null
on the
Setup
page to access the
NULL Probe Process
screen.
Fig. 21—(Left) Null probe process.
Fig. 22—(Right) The potential dierence oset shows on the main screen under the reading.
2. Place the electrodes in a saline media.
3. Push the
NULL
button. Wait for the second message stating ”Solution value 0.0”
before returning. This action removes any voltage osets on the electrodes and
neutralize the oset charge on V
1
-V
2.
4. Select the
Enable/Disable
radio button to use the oset.
5. Press
Return
. The potential dierence oset shows on the main screen under the
reading. An electrode check should show 0.0 mV. The saved data le reects the
oset.