EasyManua.ls Logo

wtw Cond 3110 - Page 20

wtw Cond 3110
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Cond 3110
20
ba75792e02 11/2008
Note
If the error message E3 appears, refer to chapter 6 W
HAT TO DO IF...
Stability control During calibration, the stability control is automatically activated.
Note
This method of automatically determining the cell constant by
calibration in the 0.01 mol/l KCL control standard solution can only be
used for measuring cells with cell constants in the range 0.450 ...
0.500 cm
-1
or 0.800 ... 0.880 cm
-1
.
3 Immerse the measuring cell in the control standard solution,
0.01 mol/l KCI.
4 Start the calibration with <ENTER>. The determination of the
cell constant with stability control starts. The display
indicator flashes until there is a stable signal.
The cell constant determined is displayed. The meter
automatically stores the cell constant.
5 Switch to the measuring mode with <ENTER>.
The determined cell constant is used.

Other manuals for wtw Cond 3110

Related product manuals