4.4 Temperature compensation TC
The calculation of the temperature compensation is based on the
preset reference temperature, Tref 20 or Tref 25 (see section
4.5 S
ETTINGS).
As the temperature compensation, the nonlinear temperature
compensation "nLF" according to DIN 38404 or EN 27 888 respectively
is permanently set.
Application ranges Test sample Temperature compensation TC Display
indicator
Natural water
(ground water,
surface water,
drinking water)
nLF
according to DIN 38404
EN 27 888
Ultrapure water nLF
according to DIN 38404
EN 27 888
Salinity
(seawater)
Automatically nLF according to
IOT
Sal,