System menu Multi 350i
26
Measuring menu:
conductivity
(continued)
System menu
The following settings and functions are available in the system menu and its
submenus:
l Viewing, transmitting and erasing stored measuring data
l Viewing and transmitting calibration records
l Display illumination and contrast
l Changing the unit of the temperature display (°C <-> °F)
l Setting the automatic switchoff
l Configuring the data interface
l Setting the menu language, date and time
l Resetting all sensor-independent system settings to the default condition
Menu item Possible
Setting
Description
Measuring cell Measuring cell used
LR325/01 LR 325/01 measuring cell,
nominal cell constant
0.100 cm
-1
.
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm
-1
.
man Any measuring cells with freely
adjustable cell constants in the
range from 0.250 to 25.000 cm
-1
.
Cell constant 0.090 to
0.110 cm
-1
Display and setting option of the
cell constant of the LR 325/01
measuring cell.
Man. cell const. 0.250 to
25.000 cm
-1
Display and setting option of the
cell constant of any measuring
cells (man).