Measuring menus Multi 350i
24
Measuring menu:
DO
Menu item Possible
setting
Description
Calibration /
Calibration interval
1 ... 999 d Calibration interval for the DO
sensor (in days).
The meter reminds you to
calibrate regularly by the flashing
sensor symbol.
Calibration /
Comparison meas.
On
Off
Enables to adjust the measured
value with the aid of a
comparison measurement, e.g.
Winkler titration.
Calibration /
Calibration record
- Displays the calibration record of
the last calibration.
Reset - Resets all sensor settings to the
delivery condition.
Sal automatic On
Off
Automatic salt content correction
for concentration
measurements.
Note:
This function is available with the
ConOx double sensor only
.
Sal correction On
Off
Manual salt content correction
for concentration
measurements.
Salinity 0.0 ... 70.0 Salinity or salinity equivalent for
the salt content correction.
This function is only available if
the manual salt content
correction is switched on.