What to do if... pH 7110
38 ba75925e01 07/2011
10 What to do if...
10.1 pH
10.1.1 No stable measured value
10.1.2 Error message CalError
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Junction of the electrode contami-
nated
Clean the junction
Glass membrane of the electrode
contaminated
Clean the glass membrane
Cause Remedy
pH value of the test sample not
stable
Measure with air excluded if
necessary
Temperature of the test sample
not stable
Adjust temperature if necessary
Cause Remedy
The values determined for zero
point and slope of the electrode
are outside the allowed limits.
Recalibrate
Junction contaminated Clean the junction
pH electrode broken Replace the pH electrode