10.2.2 Error message OFL, UFL
The measured value is outside the measuring range of the meter.
The measured value is obviously incorrect.
10.3 General information
10.3.1 Symbol for calibration evaluation flashes
10.3.2 [LoBat] display
Cause Remedy
ORP value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if necessary
Cause Remedy
Cable broken Replace the electrode
Gel electrolyte dried out Replace the electrode
Air bubble in front of the junction Remove the air bubble
Air in the junction Extract air or moisten the junction
Cause Remedy
Calibration interval expired Recalibrate the measuring system
Cause Remedy
Batteries almost empty Replace the batteries (see section
3.3.1 I
NSERTING THE BATTERIES,
page 10)